PEMF devices · professional price tier · researched 15 November 2026
OrthoCor Medical review: FDA status, field strength and what buyers report
Every fact below links to the page or record it came from. OrthoCor Medical PEMF devices were checked against FDA's own databases, field strength claims against any independent gaussmeter test, and buyer reports on Reddit, Trustpilot, BBB and Amazon. A second researcher re-checked the record, 38 items, and corrected or removed what did not hold.
- Company
- Caerus Corporation, doing business as OrthoCor Medical (510(k) holder: Caerus Corporation; FDA establishment registered as OrthoCor Medical Inc.)
- FDA status
- Holds an FDA 510(k) for its device
- Prices found
- Not published
- Website
- orthocormedical.com
Documentation score
4 of 5
Well documented
- Field strength published with units
- Frequency published
- FDA clearance or approval names its device
- Independent output test on file
- No material claim contradicted
Customer record
Not rated
Not enough reviews to rate
- Rated 4.0 or better on 50 or more reviews
- BBB rating A- or better
- No complaint repeated across 3 or more sources
Too few public reviews were found to rate this seller.
OrthoCor Medical PEMF devices on record
| Model | Type | Field strength stated | Frequency | Extras | Price | Warranty |
|---|---|---|---|---|---|---|
| Active System (wearable wraps: Shoulder, Neck, Elbow, Hand/Wrist, Back, Knee, Ankle, Foot, Universal) with single-use OrthoPods heat packs | wearable | No field strength stated on the website. FDA 510(k) summary K241395 lists Peak Generator Power 1.35 W (no magnetic field figure). The earlier K121702 predicate summary (OrthoCor Active Device, OCR of a scanned PDF) lists 'Peak B Field (T) 0.004042 T*' with the note that the value was calculated from energy density, not measured; location not stated. 0.004042 T equals 4.042 millitesla or about 40.4 gauss. | Carrier 27.12 MHz; burst frequency 2 Hz; burst width 2 ms (FDA 510(k) summary K241395). Website gives no frequency.Prescription required | heat (chemical, air-activated single-use OrthoPods) | Not published | 6 months from date of shipment, limited to replacement of defective parts, original user only |
| Activ8 System (same nine body-area wraps; built-in heating; removable rechargeable battery) | wearable | No field strength stated on the website. Not stated with units anywhere found for Activ8. | Not stated on the website. If Activ8 is the device cleared as Avenue8 under K241395, the summary lists 27.12 MHz carrier, 2 Hz burst frequency, 2 ms burst width, peak generator power 1.35 W; the Activ8 to Avenue8 link is not confirmed.Prescription required | heat (built-in electric heating per FAQ) | Not published | 6 months from date of shipment, limited to replacement of defective parts, original user only |
Is OrthoCor Medical FDA cleared?
| 510(k) or PMA | Held by | Device | Decision | Cleared or approved for | Code |
|---|---|---|---|---|---|
| K241395 | Caerus Corporation, Blaine MN | Active System; Avenue8 | 2024-12-18 | The Active System and Avenue8 are indicated for adjunctive use in the temporary relief of pain, such as pain associated with over-exertion, strains, sprains, and arthritis. | ILX (subsequent IMD, IRT) |
| K121702 | Orthocor Medical | ORTHOCOR ACTIVE DEVICE ORTHOCOR GENERAL USE ACTIVE DEVICE | 2013-05-06 | The device is indicated for adjunctive use in the palliative treatment of postoperative pain and edema in superficial soft tissue. | ILX |
| K091996 | Orthocor Medical | ORTHOCOR KNEE SYSTEM BASIC | 2009-12-15 | The OrthoCor Knee System Basic is indicated for adjunctive use in the palliative treatment of post-operative pain and edema in superficial soft tissue. | ILX |
| K092044 | Orthocor Medical | ORTHOCOR KNEE SYSTEM, MEDIAN, LARGE, X-LARGE | 2009-12-18 | Adjunctive use in the palliative treatment of postoperative edema and pain in superficial soft tissue. Temporary relief of minor aches and pains associated with overexertion, strains, sprains, and arthritis. (as quoted for the predicate in the K241395 summary table) | ILX |
FDA registration and listing
| Registered firm | Role | Products listed | Code | Country |
|---|---|---|---|---|
| ORTHOCOR MEDICAL INC. | Manufacture Medical Device | Avenue 8 | ILX, IMD, IRT (K241395) | US |
| ORTHOCOR MEDICAL INC. | Manufacture Medical Device | Orthocor | ILX (K121702) | US |
What OrthoCor Medical says about FDA, checked
- Unconfirmed “As of 2019, Active System devices are classified by the U.S. FDA as non-thermal shortwave diathermy.” About Us page. FDA cleared OrthoCor devices under the shortwave diathermy regulation 21 CFR 890.5290 (product code ILX, 'for use other than applying therapeutic deep heat') from 2009 and 2013; FDA's reclassification order renaming this type 'nonthermal shortwave therapy' was published October 13, 2015 (per the FDA cover note on K121702). The 2019 date does not match any record found. source, for the claim As of 2019, Active System devices are classified by the...
- Unconfirmed “OrthoCor devices deliver powerful relief for post-operative, acute, arthritic, and chronic pain” Home page. K241395 clears adjunctive temporary relief of pain such as from over-exertion, strains, sprains and arthritis. 'Chronic pain' as a category is broader than the cleared wording. source, for the claim OrthoCor devices deliver powerful relief for...
- Does not match “OrthoCor devices help address the root causes of pain, actively stimulating the healing cascade by increasing blood flow and reducing inflammation.” Home page. No OrthoCor clearance includes healing, blood flow or inflammation; K241395 is limited to adjunctive temporary relief of pain. source, for the claim OrthoCor devices help address the root causes of pain,...
Has OrthoCor Medical been tested?
No gaussmeter, magnetometer or oscilloscope reading of a current OrthoCor Medical device was found, from the brand or from anyone else.
- Verified “OrthoCor PST Reduces Pain and Pharmacologic Use ... A randomized, multi-site trial found that OrthoCor PST therapy significantly reduced both pain and medication use” Hackel et al., Pain Ther 2025 (PMID 39928254) tested the Orthocor Active System (27.12 MHz, 2 ms pulses at 2 Hz) vs standard of care, not sham; 91 of 120 patients analyzed; 36% vs 10% LS mean pain reduction, 42% vs 14% raw change at day 14. Conflicts: one author holds equity in Caerus Corporation through an LLC, one is a Caerus employee. Not blinded or sham controlled. source, for the claim OrthoCor PST Reduces Pain and Pharmacologic Use ... A...
- Unverified “OrthoCor Pulsed Shortwave Therapy (PST) Improves Function and Tissue Healing for Plantar Fasciitis” Indexed in PubMed: Bromley, McManus, McManus, 'Novel Approach to Plantar Fasciitis Treatment: Pulsed Electromagnetic Field Therapy', J Am Podiatr Med Assoc 2026;116(4) (PMID 42646040; seller page gives a 14 May 2025 publication date, likely online first). The abstract confirms the OrthoCor Active System was used for 12 weeks with stretching; the seller summary gives 12 patients. Uncontrolled case series (comparisons are to other studies' data), so it cannot show the device caused the improvement. source, for the claim OrthoCor Pulsed Shortwave Therapy (PST) Improves Function...
- Not supported “Effects of Pulsed Electromagnetic Fields on Interleukin-1 Beta and Postoperative Pain: A Double-Blind, Placebo-Controlled, Pilot Study in Breast Reduction Patients” Rohde et al., Plast Reconstr Surg 2010 (PMID 20527063) is a 24-patient study; the abstract does not name the device. SofPulse (Endonovo) cites the same study as evidence for its own tPEMF technology, so it is not shown to be a test of an OrthoCor device. source, for the claim Effects of Pulsed Electromagnetic Fields on Interleukin-1...
- Unverified “A meta-analysis of 11 RCTs found that PEMF therapies, including PST, significantly improved pain, stiffness, and physical function in adults with osteoarthritis.” Hu et al. 2022 meta-analysis pools various PEMF devices and signals; not a test of OrthoCor devices. source, for the claim A meta-analysis of 11 RCTs found that PEMF therapies,...
- Unverified “A review of 15 trials showed short-term benefits of PEMF therapy for knee OA” Chen et al., Physical Therapy 2020 meta-analysis of various PEMF devices; not a test of OrthoCor devices. source, for the claim A review of 15 trials showed short-term benefits of PEMF...
- Unverified “Clinically proven, drug-free pain relief ... No significant side effects” Supported for the Active System by the single company-linked open-label RCT (Hackel 2025) and the K241395 clinical study summary (36% vs 10% pain reduction, 3.3% minor adverse events). No independent trial of Activ8 found. source, for the claim Clinically proven, drug-free pain relief ... No significant...
Certifications claimed, checked
- Claim only VA Federal Supply Schedule contract FSS 36F79725D0217 (Caerus Corp d.b.a. OrthoCor Medical) Contract number stated on seller page; GSA/VA price list not retrieved. source, for the certification VA Federal Supply Schedule contract FSS 36F79725D0217...
- Certificate seen IEC 60601-1 and IEC 60601-1-2 compliance Stated in the FDA 510(k) summary K241395 performance data section. source, for the certification IEC 60601-1 and IEC 60601-1-2 compliance
OrthoCor Medical reviews: what buyers report
No Trustpilot, Amazon or BBB rating was found for OrthoCor Medical.
Repeated complaints
No complaint was reported by more than one source.
Repeated praise
No praise was repeated across sources.
OrthoCor Medical on Reddit
No Reddit thread about OrthoCor Medical could be read for this record.
Recalls, safety reports and lawsuits
| Type | Date | What the record says |
|---|---|---|
| FDA adverse event report | 2025-09-05 | Malfunction report 3008216220-2025-00001: user's battery would not hold charge; replacement battery pack got hot while charging; user said the device was hot to the touch and felt like it burned him, with no lingering redness or blistering and no medical attention needed. source, for the FDA adverse event report of 2025-09-05 |
Who sells OrthoCor Medical
- Company
- Caerus Corporation, doing business as OrthoCor Medical (510(k) holder: Caerus Corporation; FDA establishment registered as OrthoCor Medical Inc.)
- Address on record
- 8611 W 35W Service Drive NE, Suite 180, Blaine, MN 55449
- Founded
- 2007 (OrthoCor Medical, per its About page: 'In 2007, Dr. Gill founded OrthoCor Medical.')
- Owner or parent
- Caerus Corp is described on the Leadership page as 'the parent company of OrthoCor Medical'; CEO Fariborz Boor Boor is listed as founder of Caerus Corp. The VA Federal Supply Schedule entry reads 'Caerus Corp d.b.a. OrthoCor Medical'.
- Customer support
- 888.583.6268; info@orthocormedical.com; technical support support@orthocormedical.com
Sources: 1, company record2, company record3, company record4, company record5, company record6, company record
OrthoCor Medical: common questions
Is OrthoCor Medical FDA cleared?
Holds an FDA 510(k) for its device, per openFDA records checked on 15 November 2026. K241395, held by Caerus Corporation, Blaine MN, clears Active System; Avenue8 for: The Active System and Avenue8 are indicated for adjunctive use in the temporary relief of pain, such as pain associated with over-exertion, strains, sprains, and arthritis. There are 4 clearances on record in total; each is listed on this page. One FDA claim the brand makes does not match the record; they are quoted below.
Is OrthoCor Medical legit?
On the archive's documentation rule it scores 4 of 5 (well documented), and its customer record is not rated, because there are not enough reviews. None of its FDA, certification, performance or safety claims was contradicted by a record. A score measures how well the seller backs up its claims, not whether the therapy works.
What do OrthoCor Medical reviews and Reddit threads say?
The research found too few public reviews to summarise. That is stated rather than filled in.
How strong is OrthoCor Medical, and what is it FDA cleared for?
The strongest field OrthoCor Medical states is 4.042 millitesla (40.42 gauss). That is the seller's own figure; a field is strongest at the coil surface and falls off fast with distance. FDA cleared Active System; Avenue8 for: The Active System and Avenue8 are indicated for adjunctive use in the temporary relief of pain, such as pain associated with over-exertion, strains, sprains, and arthritis. It is sold on a prescription.
Has OrthoCor Medical had a recall or safety report?
The records hold 1 FDA adverse event report, each summarised and linked in the safety section. An adverse event report is an allegation filed with FDA; on its own it does not prove the device caused the event.
How these scores work
Documentation score, 5 points
One point each for: a field strength stated with its unit (gauss, tesla, millitesla or microtesla; 1 gauss is 100 microtesla); a pulse frequency stated; an FDA 510(k) clearance or premarket approval that names the seller's own device (registration and listing alone does not earn it); a gaussmeter, magnetometer or oscilloscope reading by someone other than the seller; no FDA, certification, field strength or safety claim contradicted by a record.
5 is fully documented, 4 well documented, 3 partly documented, 0 to 2 thinly documented.
Customer record, 3 points
One point each for: rated 4.0 or better on a review site with 50 or more reviews; a Better Business Bureau grade of A- or better; no single complaint reported by three or more independent buyers or sites. With too few reviews, the record says “not enough reviews to rate” instead of guessing.
Neither score says whether the therapy works. That is what the evidence pages measure.
What could not be verified
- WebSearch budget for the session was exhausted before review research; Reddit, BBB, Amazon, Google reviews, forums and YouTube were not searched. Arctic Shift was not queried for the same reason.
- Trustpilot: https://www.trustpilot.com/review/orthocormedical.com was reported as 404 by the original research; on recheck (2026-11-15) Trustpilot returned 403 (blocked), so profile existence could not be confirmed either way.
- FDA warning letter and FTC action databases were not searched (search budget exhausted); openFDA recall database shows no OrthoCor or Caerus recalls.
- No prices published; billed via Workers' Compensation and VA (HCPCS E0761 NU and A9900 NU per site). VA FSS price list for contract 36F79725D0217 not retrieved.
- Activ8 does not appear by that name in FDA 510(k) or listing records; whether Activ8 is the Avenue8 device in K241395 is unconfirmed.
- Field strength for Active System and Activ8 is not published; only a calculated 0.004042 T peak B field for the 2013 predicate device (OCR of a scanned FDA PDF).
- Seller pages are internally inconsistent on dosing: How It Works says 'Just thirty minutes of PST treatment, twice a day' while the Active System page says two hours, once per day.
- No independent lab measurement of output found.
No seller paid to be listed or rated. For what PEMF has been measured to do in people, read the PEMF evidence.