Clinic lasers and professional systems · researched 13 November 2026

LiteCure review: FDA status, lab tests and what buyers report

Every fact below links to the page or record it came from. LiteCure devices were checked against FDA's own databases, brightness claims against independent optical tests, and buyer reports on Reddit, Trustpilot, BBB and Amazon. A second researcher re-checked the record, 41 items, and corrected or removed what did not hold.

Company
LiteCure, LLC
FDA status
Holds an FDA 510(k) for its device
Prices found
Not published

Documentation score

2 of 5

Thinly documented

  • Wavelengths published
  • Brightness published with its distance
  • Optical test on file
  • Traceable at FDA
  • No material claim contradicted

Customer record

2 of 3

Mixed customer record

  • Rated 4.0 or better on 50 or more reviews
  • BBB rating A- or better
  • No complaint repeated across 3 or more sources

LiteCure devices on record

As listed by the seller on 13 November 2026. Specifications are the seller's own figures unless a test below says otherwise.
ModelTypeWavelengths (nm)Brightness claimedPriceWarranty
LightForce XLi (40W) Therapy Laser System (model LTS-4000-L-S8 per user manual)laser810, 980Not publishedNot published24 months
LightForce XPi (25W) Therapy Laser System (model LTS-2500-L-S8 per user manual)laser810, 980Not publishedNot published24 months
LightForce FXi (15W) Therapy Laser (model LTC-1510-L-B6 per user manual)laser810, 980Not publishedNot published24 months
Companion CT-RevitL Therapy Laser (veterinary)laser808, 980, 1064Not publishedNot publishedNot stated
Companion CTX-IQ Therapy Laser (veterinary)laser810, 980Not publishedNot published4 years
Companion CTX-15 Therapy Laser (veterinary)laserNot listedNot publishedNot published3-year standard warranty
Companion CTX-10 Therapy Laser (veterinary)laser808, 980Not publishedNot published2-year standard warranty for basic console coverage, 90 days for handpiece and accessories
Pegasus PTX-IQ Therapy Laser (equine)laser810, 980Not publishedNot publishedNot stated

Is LiteCure FDA cleared?

Holds an FDA 510(k) for its device. LiteCure, LLC holds K173067 (LightForce LTS Model 1000, 1500, 2500 and 4000). The current LightForce user manual identifies the XPi as LTS-2500-L-S8 and the XLi as LTS-4000-L-S8, so those two are covered. The FXi manual identifies it as LTC-1510-L-B6; no 510(k) naming LTC-1510 or FXi was found in openFDA (K121487 names LTC-900 and LTC-1200 only). No 510(k) was found for Companion or Pegasus veterinary model names. These are 810/980 nm near-infrared Class IV lasers (650 nm aiming beam), not red LED devices; cleared use is prescription, professional only. A 510(k) clears one named device for one stated use. Registration and listing is a filing by the company, not an FDA review.
510(k)Held byCleared deviceDecisionCleared forCode
K173067Litecure, LLCLightForce LTS Model 1000, 1500, 2500, and 40002018-02-22810 nm and 980 nm wavelength: The LTS-1000/1500/2500/4000 devices emit energy in the visible and near infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and the temporary relaxation of muscle. (The LTS-2500/4000 are also cleared at 980 nm for listed surgical applications and for temporary increase of clear nail in onychomycosis.)GEX
K103511Litecure, LLCLITECURE THERAPY SYSTEM2011-01-25LiteCure Therapy System, Model LTS-1500 is indicated for emitting energy in the infrared Spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis and promoting relaxation of the muscle tissue and to temporarily increase local blood circulation.ILY
K121487Litecure, LLCLITECURE THERAPY COMPACT2012-06-12LiteCure Therapy Compact (Model LTC-900 and LTC-1200) is indicated for emitting energy in the infrared Spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis and promoting relaxation of the muscle tissue and to temporarily increase local blood circulation.ILY
K123014Litecure, LLCLITECURE THERAPY SYSTEM MODEL LTS-15002013-03-29Summary not readablePDZ
K093087Litecure, LLCBWF-5 MEDICAL LASER SERIES2010-05-12Summary not readableGEX
K123031Litecure, LLCBWF-5 MEDICAL LASER SERIES2012-12-03Summary not readableGEX
K070400Litecure, LLCLC THERAPY, MODEL LCT-10002007-02-28Summary not readableILY

FDA registration and listing

Registered firmRoleProducts listedCodeCountry
LiteCure, LLCManufacture Medical DeviceGeneral Surgical Laser, Medical Laser SeriesILY, GEX, PDZUS
LT Technology Limited (owner/operator LITECURE, LLC)Manufacture Medical DeviceILY, PDZ, GEXCN

What LiteCure says about FDA, checked

  • Unconfirmed “USFDA Medical Device: Class II Medical Device per USFDA 21CFR 890.5500” LightForce FXi user manual specifications table. 21 CFR 890.5500 (product code ILY) is Class II in FDA's classification data, so the classification is consistent, but no 510(k) naming the FXi or model LTC-1510 was found. source, for the claim USFDA Medical Device: Class II Medical Device per USFDA...
  • Matches “The laser system emits energy in the visible and near-infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; the temporary relaxation of muscle.” LightForce XLi/XPi user manual, Indications for Use. Matches the K173067 indication for LTS-2500 and LTS-4000, the model numbers the manual assigns to XPi and XLi. source, for the claim The laser system emits energy in the visible and...
  • Does not match “Indications: Deep Tissue Laser Therapy is used to treat painful muscles and joints associated with acute and chronic soft tissue injuries. It can also relieve pain and stiffness associated with arthritis and is able to increase localized blood flow.” Chattanooga Rehab LightForce XLi, XPi and FXi product pages. Broader wording than the cleared K173067 indication, which is limited to topical heating for temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm, temporary increase in local blood circulation and temporary muscle relaxation. source, for the claim Indications: Deep Tissue Laser Therapy is used to treat...
  • Unconfirmed “The FXi complies with 21 CFR Chapter 1, Sub chapter J, as administered by the Center for Devices and Radiological Health (CDRH) of the US Food and Drug Administration (FDA).” LightForce FXi user manual. Radiation-product compliance is not checkable through the openFDA endpoints used. source, for the claim The FXi complies with 21 CFR Chapter 1, Sub chapter J, as...

Has LiteCure been lab tested?

Only a spectroradiometer or an integrating sphere reads LED output reliably; solar power meters over-read. A brand figure is only compared with a test taken at the same distance. Who paid for the test, and whether the tester earns from links, is printed beside it.

No optical test of a current LiteCure device was found, from the brand or from anyone else.

  • Unverified “Clinically Studied to Help in the Treatment of: Neck Pain, Low Back Pain, Shoulder Pain, Elbow Pain, TMJ, Arthritis, Plantar Fasciitis, Sports Injuries” The ten cited references are general low-level and high-intensity laser therapy literature (for example Chow et al., Lancet 2009 meta-analysis). The page does not state that any cited study used a LightForce device. No independent output measurement of any LiteCure laser was found. source, for the claim Clinically Studied to Help in the Treatment of: Neck Pain,...

Certifications claimed, checked

UL, ETL and IEC 60601 are electrical safety marks. ISO 13485 is a quality system, not a test of the device's light output.

LiteCure reviews: what buyers report

Read from public pages on 13 November 2026. A complaint counts as repeated when three or more independent buyers or sites report it.
WhereScoreReviewsNote
Better Business BureauA+0 complaints in 3 years

Repeated complaints

No complaint was reported by more than one source.

Repeated praise

No praise was repeated across sources.

LiteCure on Reddit

No Reddit thread about LiteCure could be read for this record.

Claims that did not match the record

Each claim is quoted from the seller's own pages and set against the record that contradicts it. FDA, certification, performance and safety claims count against the documentation score. Marketing numbers that do not match are listed but not scored.
  • Safety “Indications: Deep Tissue Laser Therapy is used to treat painful muscles and joints associated with acute and chronic soft tissue injuries. (Chattanooga Rehab LightForce XLi, XPi and FXi product pages)”Record: K173067 clears the LTS-2500/4000 (XPi/XLi) for topical heating for temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain or muscle spasm, temporary increase in local blood circulation and temporary relaxation of muscle (plus 980 nm surgical and onychomycosis indications). It does not clear treatment of acute or chronic soft tissue injuries. The FXi has no 510(k) found at all. source, for the record against Indications: Deep Tissue Laser Therapy is used to treat...
  • Marketing “In 2021, DJO acquired LiteCure. (Chattanooga Rehab LightForce page, chattanoogarehab.com/us/lightforcemedical)”Record: LiteCure's own newswire page carries the DJO press release dated December 11, 2020 announcing the acquisition, and the Companion Animal Health About page says the acquisition was in 2020. source, for the record against In 2021, DJO acquired LiteCure. (Chattanooga Rehab...

Recalls, safety reports and lawsuits

From FDA's recall, adverse event and warning letter records, and court or regulator filings. An adverse event report is an allegation filed with FDA; on its own it does not prove the device caused the event.
TypeDateWhat the record says
FDA adverse event report2024-08-15Voluntary report (MW5158524), LightForce therapy laser: patient reported severe upper arm pain within 20 minutes of deep tissue laser therapy to left shoulder and arm, then a palm sore and blistering down to the fingers within 30 hours. Event type Injury. source, for the FDA adverse event report of 2024-08-15
FDA adverse event report2024-10-04Manufacturer report 9616086-2024-00089, LightForce XLi (40W): same shoulder and arm blistering event as above, reported by the manufacturer. Event type Injury. source, for the FDA adverse event report of 2024-10-04
FDA adverse event report2025-10-01Manufacturer report 9616086-2025-00063, LightForce XLi (40W): returned unit confirmed no aiming beam and hot emission port; shorted emission port, badly burned handpiece tip, shorted 810 nm and 650 nm diodes; parts replaced. Event type Injury. source, for the FDA adverse event report of 2025-10-01
FDA adverse event report2025-12-16Manufacturer report 9616086-2025-00075, Companion LTS-2501-C-S8 (CTS-DUO), veterinary: skin abnormalities after treatment later recognized as burns. Device not returned; root cause not established. source, for the FDA adverse event report of 2025-12-16
FDA adverse event report2025-12-17Manufacturer report 9616086-2025-00074, Companion LTS-2501-C-S8 (CTS-DUO): reported that a tech burned a patient. Device not returned; root cause not established. source, for the FDA adverse event report of 2025-12-17
FDA adverse event report2026-02-11Manufacturer report 9616086-2026-00003, Companion LTS-2501-C-S8 (CTS-DUO): user reported treatment heads overheating; a user testing a head on themselves got a burn blister. Root cause not established. source, for the FDA adverse event report of 2026-02-11
FDA adverse event report2026-03-07Manufacturer report 9616086-2026-00006, LightForce XLi (40W): customer reported handpiece randomly becoming extremely hot; two employees sustained blisters on the thumb. Event type Injury. source, for the FDA adverse event report of 2026-03-07
FDA adverse event report2026-03-18Manufacturer report 9616086-2026-00007, Companion PTX-IQ (equine): customer reported the laser had been burning patients and that two staff members burned themselves testing it on their own skin. Unit was later returned and tested; output within specification; complaint not confirmed. Event type Injury. source, for the FDA adverse event report of 2026-03-18
FDA adverse event report2026-04-29Manufacturer report 9616086-2026-00011, LTS-2501-C-S8 (CTS-DUO): customer reported a patient thermal burn; no further information; device not returned. source, for the FDA adverse event report of 2026-04-29
FDA adverse event report2026-04-29Manufacturer report 9616086-2026-00012, LTS-2501-C-S8 (CTS-DUO): customer reported a patient's fur singed and skin burned; device not returned. source, for the FDA adverse event report of 2026-04-29
FDA adverse event report2026-04-29Manufacturer report 9616086-2026-00014, LTS-2501-C-S8 (CTS-DUO): pet owner reported a dog burned with skin sloughing after laser treatment at a hospital; device not returned. source, for the FDA adverse event report of 2026-04-29
FDA adverse event report2026-04-29Manufacturer report 9616086-2026-00010, Companion CTX-IQ: burn-like lesions on the back of a Pomeranian dog treated for cervical arthritis; device not returned. source, for the FDA adverse event report of 2026-04-29
FDA adverse event report2026-05-06Manufacturer report 9616086-2026-00016, LightForce XLi (40W): redness noted on a human patient during treatment; patient later reported a third degree burn. source, for the FDA adverse event report of 2026-05-06
FDA adverse event report2017-02-01Voluntary report MW5067628, LiteCure LCT-1000 therapeutic laser: cable at the unit connector caught fire and melted due to a short; treatment stopped; report states no patient injury (openFDA event type field reads Injury). source, for the FDA adverse event report of 2017-02-01
FDA adverse event report2017-02-09Voluntary report MW5067844, LiteCure LCT-1000 class IV laser: patient reported heart pounding, dizziness, headache and nausea during and after lower back laser treatment. source, for the FDA adverse event report of 2017-02-09
FDA adverse event report2017-11-10Manufacturer report 3006268867-2017-00001, LightForce therapy laser: injury event; manufacturer noted the trained staff had left the practice and the manual warns against untrained use and use with poor circulation. source, for the FDA adverse event report of 2017-11-10
FDA adverse event report2019-03-20Manufacturer report 3006268867-2019-00002, LightForce therapy laser: injury report; no product malfunction suspected; device returned for evaluation. source, for the FDA adverse event report of 2019-03-20
FDA adverse event report2019-03-20Manufacturer report 3006268867-2019-00001, LightForce therapy laser: device evaluated, no quality issues found; injury attributed by manufacturer to a known risk documented in labeling. source, for the FDA adverse event report of 2019-03-20

Who sells LiteCure

As found in the seller's own terms and contact pages, BBB and FDA registration, with the source for each.
Company
LiteCure, LLC
Address on record
litecure.com footer: 101 Lukens Dr, Ste A, New Castle, DE 19720. FDA establishment registration 3026668625 lists 3300 Eagle Parkway, Fort Worth, TX. BBB and the K173067 510(k) summary list 250 Corporate Blvd, Suite B, Newark, DE 19702.
Founded
2006 (BBB lists business started 1/1/2006; Companion About page says laser physicist Brian Pryor, PhD, founded Companion in 2006)
Owner or parent
DJO, LLC announced its acquisition of LiteCure on December 11, 2020. DJO became Enovis in 2022 (per Chattanooga Rehab page). LightForce is now sold under Chattanooga, a brand of DJO, LLC, an Enovis company; Companion Animal Health says it was acquired by DJO, now Enovis (NYSE: ENOV), in 2020.
Customer support
877.627.3858 (litecure.com footer). LightForce sales and service now run through Chattanooga Rehab (chattanoogarehab.com/us/customer-service-lightforce).

Sources: 1, company record2, company record3, company record4, company record5, company record6, company record

LiteCure: common questions

Is LiteCure FDA cleared?

Holds an FDA 510(k) for its device, per openFDA records checked on 13 November 2026. K173067, held by Litecure, LLC, clears LightForce LTS Model 1000, 1500, 2500, and 4000 for: 810 nm and 980 nm wavelength: The LTS-1000/1500/2500/4000 devices emit energy in the visible and near infrared spectrum to provide topical heating for the purpose of elevating. One FDA claim the brand makes does not match the record; they are quoted below.

Is LiteCure legit?

LiteCure is sold by LiteCure, LLC. On the archive's documentation rule it scores 2 of 5 (thinly documented), and its customer record is 2 of 3 (mixed customer record). 1 of its FDA, certification, performance or safety claims did not match the record. A score measures how well the seller backs up its claims, not whether the therapy works.

What do LiteCure reviews and Reddit threads say?

The research found too few public reviews to summarise. That is stated rather than filled in.

What wavelengths do LiteCure devices use?

Across its current devices the brand lists 4 distinct wavelengths: 808 nm, 810 nm, 980 nm and 1064 nm. Listed wavelengths are the seller's figures; where an independent test checked the peaks, it is shown in the lab test table.

Has anyone lab tested LiteCure?

No spectroradiometer or integrating sphere test of a current device was found, from the brand or anyone else. Brightness figures for this brand are the seller's own.

Has LiteCure had a recall or safety report?

The records hold 18 FDA adverse event reports, each summarised and linked in the safety section. An adverse event report is an allegation filed with FDA; on its own it does not prove the device caused the event.

How these scores work

Documentation score, 5 points

One point each for: wavelengths published; brightness published with the distance it was measured at; a spectroradiometer or integrating sphere test of a current device on file; traceable at FDA through a 510(k) or a registration and listing; no FDA, certification, performance or safety claim contradicted by a record or by a test at the seller's own distance.

5 is fully documented, 4 well documented, 3 partly documented, 0 to 2 thinly documented.

Customer record, 3 points

One point each for: rated 4.0 or better on a review site with 50 or more reviews; a Better Business Bureau grade of A- or better; no single complaint reported by three or more independent buyers or sites. With too few reviews, the record says “not enough reviews to rate” instead of guessing.

Neither score says whether the therapy works. That is what the evidence pages measure.

What could not be verified

  • Reddit, YouTube, forum and general review searches were not completed: the shared web search budget for this run was exhausted before LiteCure searches ran. The Arctic Shift Reddit archive was not used because the run's search limit barred archive-service searching as a workaround. Reddit threads_read is 0.
  • No Trustpilot profile exists at trustpilot.com/review/litecure.com (HTTP 404). Chattanooga and Companion profiles not checked.
  • No 510(k) naming the LightForce FXi (LTC-1510-L-B6) was found; whether it is covered by an existing clearance is unresolved.
  • No 510(k) found for Companion or Pegasus veterinary laser model names; their US regulatory basis was not established.
  • Prices are quote only (demo request); no list prices published.
  • No independent irradiance or power measurement of any LiteCure, LightForce or Companion laser was found.
  • FDA warning letter database was not searched beyond openFDA device endpoints.
  • Role of LT Technology Limited (China establishment listed under LiteCure, LLC) not described on any LiteCure page read.

No seller paid to be listed or rated. For what red light has been measured to do in people, read the red light evidence.