PEMF devices · professional price tier · researched 15 November 2026

Regenesis Biomedical review: FDA status, field strength and what buyers report

Every fact below links to the page or record it came from. Regenesis Biomedical PEMF devices were checked against FDA's own databases, field strength claims against any independent gaussmeter test, and buyer reports on Reddit, Trustpilot, BBB and Amazon. A second researcher re-checked the record, 42 items, and corrected or removed what did not hold.

Company
Regenesis Biomedical, Inc.
FDA status
Holds an FDA 510(k) for its device
Prices found
$5,528.50 to $10,081.50

Documentation score

3 of 5

Partly documented

  • Field strength published with units
  • Frequency published
  • FDA clearance or approval names its device
  • Independent output test on file
  • No material claim contradicted

Customer record

Not rated

Not enough reviews to rate

  • Rated 4.0 or better on 50 or more reviews
  • BBB rating A- or better
  • No complaint repeated across 3 or more sources

Too few public reviews were found to rate this seller.

Regenesis Biomedical PEMF devices on record

As listed by the seller on 15 November 2026. Specifications are the seller's own figures unless a test below says otherwise.
ModelTypeField strength statedFrequencyExtrasPriceWarranty
Reprieve by Regenesis (two treatment applicator pads), part 421-2000-02local applicatorNo magnetic field strength stated. 510(k) summary K223620: RF power less than 10 W, pulsed mode output 0.3 to 7.0 W; achieves 4 degrees C temperature rise at 25 minutes (deep heat device).27.12 MHz carrier; continuous wave mode or pulsed mode with pulse rate 200 Hz to 1000 Hz in 200 Hz steps and pulse width 20 to 100 microseconds (K223620 summary)Prescription requireddeep heat (thermal shortwave diathermy)$9,989.5Repair or replacement of a defective product throughout the evaluation period or within one year for a purchased product; applies only to product used in the United States (VA Federal Supply Schedule price list terms).
Reprieve by Regenesis CW (one treatment applicator pad), part 421-2000-00local applicatorNo magnetic field strength stated.27.12 MHz (K223620); CW model name suggests continuous wave, not stated in detail on the sitePrescription requireddeep heat$10,081.5Repair or replacement of a defective product throughout the evaluation period or within one year for a purchased product; applies only to product used in the United States (VA Federal Supply Schedule price list terms).
Reprieve 60 by Regenesis (Compliance and Compatibility Assessment, two pads, active 60 days), part 421-2000-62local applicatorNo magnetic field strength stated.27.12 MHz (K223620)Prescription requiredNone stated$5,528.5Repair or replacement of a defective product throughout the evaluation period or within one year for a purchased product; applies only to product used in the United States (VA Federal Supply Schedule price list terms).
Provant Therapy System, Model 4201 / Provant Infinity (manual linked in site footer; not listed on Products page)local applicatorElectrical field strength of 591 V/m at 5 cm above the Treatment Applicator; average RF generator power 1.9 Watts (K131979 summary, OCR of scanned PDF). No magnetic field figure.27.12 MHz carrier, pulsed at 1 kHz with 42 microsecond pulse width, 4% duty cycle (K131979 summary)Prescription requiredNone statedNot publishedNot stated

Is Regenesis Biomedical FDA cleared?

Holds an FDA 510(k) for its device. Regenesis holds its own clearances. Note the current product, Reprieve, was cleared in 2023 under K223620 as a thermal shortwave diathermy device (product code IMJ, 'to generate deep heating within body tissues'), not as a non-thermal PEMF device. The older Provant clearances (K091791, K131979) are non-thermal ILX clearances for post-operative pain and edema. A 510(k) clears one named device for one stated use; a premarket approval (PMA) approves one after FDA reviews clinical evidence, the route bone growth stimulators take. Registration and listing is a filing by the company, not an FDA review.
510(k) or PMAHeld byDeviceDecisionCleared or approved forCode
K223620Regenesis Biomedical, Inc.Reprieve by Regenesis2023-05-24The Reprieve by Regenesis™ device is indicated to generate deep heating within body tissues for the treatment of conditions such as relief of pain and muscle spasms.IMJ
K131979Regenesis Biomedical, Inc.PROVANT THERAPY SYSTEM2013-12-13The Provant Therapy System is indicated for adjunctive use in the palliative treatment of postoperative pain and edema in superficial soft tissue.ILX
K091791Regenesis Biomedical, Inc.PROVANT SYSTEM, MODEL 42012010-04-07The Provant System is intended for adjunctive use in the palliative treatment of postoperative pain and edema in superficial soft tissue.ILX
K972093Regenesis Biomedical, Inc.REGENESIS MODEL 421997-10-21Summary not readableILX

FDA registration and listing

Registered firmRoleProducts listedCodeCountry
REGENESIS BIOMEDICAL, INC.Develop Specifications But Do Not Manufacture At This FacilityReprieve by RegenesisIMJUS
REGENESIS BIOMEDICAL, INC.Develop Specifications But Do Not Manufacture At This FacilityProvant Therapy SystemILXUS
Velentium MedicalManufacture Medical Device for Another Party (Contract Manufacturer)Reprieve by RegenesisIMJUS

What Regenesis Biomedical says about FDA, checked

  • Matches “PRESCRIPTION BASED, FDA CLEARED” For Patients page. K223620 is a prescription clearance for Reprieve. source, for the claim PRESCRIPTION BASED, FDA CLEARED
  • Matches “it has been specifically designed and FDA-cleared for safe use at home.” For Patients page. K223620 summary describes a prescription device designed for home use by a nontechnical operator. source, for the claim it has been specifically designed and FDA-cleared for safe...
  • Unconfirmed “Reprieve by Regenesis ... an electromagnetic shortwave diathermy therapy (with conditions listed including diabetic peripheral neuropathy pain and limb amputation (phantom limb) pain)” Home page ('an electromagnetic shortwave diathermy therapy'); the diabetic peripheral neuropathy and phantom limb pain wording appears on the For Patients page only. Cleared indication is deep heating for conditions 'such as relief of pain and muscle spasms'. Neuropathy and phantom limb pain are not named in the clearance; the DPN trial the site cites used Provant-type pulsed settings (1000 Hz, 42 microseconds), not a deep-heat protocol. source, for the claim Reprieve by Regenesis ... an electromagnetic shortwave...

Has Regenesis Biomedical been tested?

A gaussmeter or magnetometer reads the field a coil puts out, and an oscilloscope reads its pulse shape and frequency. The field is strongest at the coil surface and falls off fast with distance, so a reading means little without where it was taken. Who paid for the test, and whether the tester earns from links, is printed beside it.

No gaussmeter, magnetometer or oscilloscope reading of a current Regenesis Biomedical device was found, from the brand or from anyone else.

  • Verified “Diabetic Peripheral Neuropathy Clinical Trial Results” Tassone, Page, Slepian, J Diabetes Sci Technol 2025 (PMID 37542366): 182 subjects, sham controlled; a 30% pain reduction vs sham only in active patients who did not receive a low field strength notification (a subgroup). Device: Regenesis 27.12 MHz, 42 microsecond pulses at 1000 Hz, 4.2% duty cycle. Funded by Regenesis; first author is a Regenesis employee, others on its advisory board and board of directors. source, for the claim Diabetic Peripheral Neuropathy Clinical Trial Results
  • Verified “Lumbar Clinical Trial Results” Sorrell et al., J Pain Res 2018 (PMID 29950893): 36-subject pilot, 42 microsecond arm (the Provant Therapy System setting, per the full text) vs 38 microsecond arm vs sham; low-back pain fell 40.2% in the 42 microsecond arm, and the authors conclude only 'trends' versus sham. Two of five authors list Regenesis Biomedical as their affiliation, although the paper's conflict statement reports no conflicts. Tested Provant, not Reprieve. source, for the claim Lumbar Clinical Trial Results
  • Unverified “Retrospective Study for Musculoskeletal Pain (Marovino 2023, PEMF therapy)” Retrospective design; full paper not reviewed; device used not confirmed. source, for the claim Retrospective Study for Musculoskeletal Pain (Marovino...
  • Unverified “Over 30,000 patients treated” Company figure; no independent source. The 2013 K131979 summary cites complaint data from over 175,000 patients and over 3,000,000 treatments (industry wide, including other firms). source, for the claim Over 30,000 patients treated

Certifications claimed, checked

UL, ETL and IEC 60601 are electrical safety marks. ISO 13485 is a quality system. None of them measures the field a device puts out.

Regenesis Biomedical reviews: what buyers report

Read from public pages on 15 November 2026. A complaint counts as repeated when three or more independent buyers or sites report it.

No Trustpilot, Amazon or BBB rating was found for Regenesis Biomedical.

Repeated complaints

No complaint was reported by more than one source.

Repeated praise

No praise was repeated across sources.

Regenesis Biomedical on Reddit

No Reddit thread about Regenesis Biomedical could be read for this record.

Recalls, safety reports and lawsuits

From FDA's recall, adverse event and warning letter records, and court or regulator filings. An adverse event report is an allegation filed with FDA; on its own it does not prove the device caused the event.
TypeDateWhat the record says
Recall2008-09-29Recall Z-0887-2009 (event 50230), PROVANT Wound Therapy System Model 4201: firm found affected devices may, under certain unusual clinical circumstances, be unable to generate a therapeutic dose of RF energy. 7 units; status Terminated 2009-07-08. source, for the Recall of 2008-09-29
FDA adverse event report2008-06-052032108-2008-00001 Injury: patient treated for bilateral lower leg ischemic ulcers with Provant plus Panafil reported a progressive burning sensation at the right ankle ulcer during sessions; both treatments were stopped, and two days later the ER found a small area of blistering and necrosis next to the wound suggestive of a third degree burn, which resolved with debridement and antibiotics. Manufacturer found the device within specification; cause unknown. source, for the FDA adverse event report of 2008-06-05
FDA adverse event report2008-09-182032108-2008-00002 Injury report: squamous cell carcinoma diagnosed 10 months after Provant therapy in a chronic pressure ulcer; clinician concluded Provant did not cause or contribute. source, for the FDA adverse event report of 2008-09-18
FDA adverse event report2009-03-232032108-2009-00001 Injury: paraplegic patient with sacral stage IV pressure ulcer treated with Provant and Panafil developed several asymptomatic 2 cm blisters on day 5; Provant was stopped, then resumed, and skin erosions consistent with de-roofed blisters appeared six days later. Clinician's diagnosis: peri-wound vesicular skin reaction, possibly due to Provant or possibly a hypersensitivity reaction to Panafil. source, for the FDA adverse event report of 2009-03-23
FDA adverse event report2011-07-132032108-2011-00001 Injury report: surveillance biopsy of a 10 year old suspected venous leg ulcer, treated with Provant over 41 discontinuous months, found focal squamous cell carcinoma (Marjolin's ulcer). Prescribing clinician did not consider Provant a cause; device returned and found within specification. source, for the FDA adverse event report of 2011-07-13

Who sells Regenesis Biomedical

As found in the seller's own terms and contact pages, BBB and FDA registration, with the source for each.
Company
Regenesis Biomedical, Inc.
Address on record
5301 N. Pima Road, Suite 150, Scottsdale, AZ 85250
Founded
1996 (About page: 'founded in 1996 on a technology platform conceived and developed by Dr. Frank George and Dr. Mary Ritz')
Owner or parent
Private; no parent named. CEO Tom Eisiminger, Jr.; Chairman Richard D. Shirk. VA FSS price list lists it as a small business.
Customer support
(877) 970-4970; info@regenesismed.com; Patient Care Coordinators

Sources: 1, company record2, company record3, company record4, company record5, company record6, company record

Regenesis Biomedical: common questions

Is Regenesis Biomedical FDA cleared?

Holds an FDA 510(k) for its device, per openFDA records checked on 15 November 2026. K223620, held by Regenesis Biomedical, Inc., clears Reprieve by Regenesis for: The Reprieve by Regenesis™ device is indicated to generate deep heating within body tissues for the treatment of conditions such as relief of pain and muscle spasms. There are 4 clearances on record in total; each is listed on this page.

Is Regenesis Biomedical legit?

Regenesis Biomedical is sold by Regenesis Biomedical, Inc. On the archive's documentation rule it scores 3 of 5 (partly documented), and its customer record is not rated, because there are not enough reviews. None of its FDA, certification, performance or safety claims was contradicted by a record. A score measures how well the seller backs up its claims, not whether the therapy works.

What do Regenesis Biomedical reviews and Reddit threads say?

The research found too few public reviews to summarise. That is stated rather than filled in.

How strong is Regenesis Biomedical, and what is it FDA cleared for?

Regenesis Biomedical does not state a field strength with a unit the research could read. FDA cleared Reprieve by Regenesis for: The Reprieve by Regenesis™ device is indicated to generate deep heating within body tissues for the treatment of conditions such as relief of pain and muscle spasms. It is sold on a prescription.

Has Regenesis Biomedical had a recall or safety report?

The records hold 1 recall and 4 FDA adverse event reports, each summarised and linked in the safety section. An adverse event report is an allegation filed with FDA; on its own it does not prove the device caused the event.

How these scores work

Documentation score, 5 points

One point each for: a field strength stated with its unit (gauss, tesla, millitesla or microtesla; 1 gauss is 100 microtesla); a pulse frequency stated; an FDA 510(k) clearance or premarket approval that names the seller's own device (registration and listing alone does not earn it); a gaussmeter, magnetometer or oscilloscope reading by someone other than the seller; no FDA, certification, field strength or safety claim contradicted by a record.

5 is fully documented, 4 well documented, 3 partly documented, 0 to 2 thinly documented.

Customer record, 3 points

One point each for: rated 4.0 or better on a review site with 50 or more reviews; a Better Business Bureau grade of A- or better; no single complaint reported by three or more independent buyers or sites. With too few reviews, the record says “not enough reviews to rate” instead of guessing.

Neither score says whether the therapy works. That is what the evidence pages measure.

What could not be verified

  • regenesismed.com blocks curl (HTTP/2 protocol error); pages were read via WebFetch. The Provant Infinity manual PDF exceeded the fetch size limit and was not read; whether Provant is still sold is unclear (Products page lists only Reprieve models).
  • WebSearch budget for the session was exhausted; Reddit, BBB, Amazon, Google reviews, forums and YouTube were not searched; Arctic Shift not queried.
  • Trustpilot: https://www.trustpilot.com/review/regenesismed.com was reported as 404 by the original research; on recheck (2026-11-15) Trustpilot returned 403 (blocked), so profile existence could not be confirmed either way.
  • FDA warning letter and FTC databases not searched; openFDA shows one terminated 2008 Provant recall and no Reprieve MAUDE reports.
  • Prices are VA Federal Supply Schedule prices, not consumer prices; no consumer price list published.
  • No magnetic field strength published for any model; no independent lab test found.
  • Patient Testimonials page not reviewed.

No seller paid to be listed or rated. For what PEMF has been measured to do in people, read the PEMF evidence.