PEMF devices · professional price tier · researched 15 November 2026

SofPulse review: FDA status, field strength and what buyers report

Every fact below links to the page or record it came from. SofPulse PEMF devices were checked against FDA's own databases, field strength claims against any independent gaussmeter test, and buyer reports on Reddit, Trustpilot, BBB and Amazon. A second researcher re-checked the record, 34 items, and corrected or removed what did not hold.

Company
Endonovo Therapeutics, Inc. (Delaware)
FDA status
Holds an FDA 510(k) for its device
Prices found
Not published

Documentation score

2 of 5

Thinly documented

  • Field strength published with units
  • Frequency published
  • FDA clearance or approval names its device
  • Independent output test on file
  • No material claim contradicted

Customer record

Not rated

Not enough reviews to rate

  • Rated 4.0 or better on 50 or more reviews
  • BBB rating A- or better
  • No complaint repeated across 3 or more sources

Too few public reviews were found to rate this seller.

SofPulse PEMF devices on record

As listed by the seller on 15 November 2026. Specifications are the seller's own figures unless a test below says otherwise.
ModelTypeField strength statedFrequencyExtrasPriceWarranty
SofPulse targeted PEMF (tPEMF) disposable wearable applicator (FDA-listed model family: SofPulse 912-M10, Roma3, Torino II)local applicatorNo field strength in gauss or tesla stated. Seller: emits 'a pulsed electromagnetic field that has 1,000 times LESS power than your mobile phone' (no units). FDA 510(k) summary K070541 states energy deposited per pulse 6.5 microwatt-seconds per cubic cm (OCR of scanned PDF) and mean duty cycle 2.5%; no magnetic field figure.27.12 MHz RF carrier (K070541 summary; Endonovo 10-K: 'pulsed short-wave radiofrequency at 27.12 MHz'). Pulse repetition rate not stated on site.Prescription requiredNone statedNot publishedNot stated

Is SofPulse FDA cleared?

Holds an FDA 510(k) for its device. K070541 (2008) was cleared to Ivivi Technologies, Inc., not to Endonovo. Endonovo acquired the SofPulse assets in 2017 (10-K) and holds an active FDA listing (FEI 3014308816) for Torino, Roma, Sofpulse, WoundPulse and OrthoHalo citing K070541, so the clearance appears to have been transferred with the assets; no transfer record was retrieved. Cleared use is limited to adjunctive palliative treatment of post-operative pain and edema in superficial soft tissue, prescription use. Wound healing is not in the cleared indication. A 510(k) clears one named device for one stated use; a premarket approval (PMA) approves one after FDA reviews clinical evidence, the route bone growth stimulators take. Registration and listing is a filing by the company, not an FDA review.
510(k) or PMAHeld byDeviceDecisionCleared or approved forCode
K070541Ivivi Technologies, Inc., Northvale NJ (original applicant)IVIVI SOFPULSE, MODELS 912-M10, ROMA3 AND TORINO II2008-12-11Adjunctive use in the palliative treatment of post-operative pain and edema in superficial soft tissue.ILX

FDA registration and listing

Registered firmRoleProducts listedCodeCountry
ENDONOVO THERAPEUTICSDevelop Specifications But Do Not Manufacture At This FacilityTorino, Roma, Sofpulse, WoundPulse, OrthoHaloILXUS
ENDONOVO THERAPEUTICSDevelop Specifications But Do Not Manufacture At This FacilityZeobiILX (K121338, held by Ivivi Health Science, LLC)US
ADM TRONICS UNLTD, INC.Manufacture Medical Device for Another Party (Contract Manufacturer)SofPulse, Roma, TorinoILXUS

What SofPulse says about FDA, checked

  • Matches “FDA Cleared for treating post-operative pain and edema in soft tissue” Technology page. Matches K070541 (adjunctive palliative treatment of post-operative pain and edema in superficial soft tissue), although the clearance was issued to Ivivi Technologies. source, for the claim FDA Cleared for treating post-operative pain and edema in...
  • Does not match “Endonovo tPEMF technology is FDA approved for the treatment of post-operative pain and edema.” Pain page. The device has 510(k) clearance (substantial equivalence), not FDA approval; no PMA found. source, for the claim Endonovo tPEMF technology is FDA approved for the treatment...
  • Does not match “SofPulse has CMS national coverage for wound healing and is supported by extensive clinical studies to treat acute, chronic and diabetic wounds” Wound Healing page. Wound healing is not in the K070541 cleared indication. CMS NCD 270.1 covers electromagnetic therapy as a category for chronic Stage III/IV pressure ulcers and arterial, diabetic and venous stasis ulcers after 30 days of failed standard care; it does not name SofPulse and does not cover 'acute' wounds. source, for the claim SofPulse has CMS national coverage for wound healing and is...
  • Does not match “These statements have not been evaluated by the Food and Drug Administration. SofPulse is not intended to diagnose, treat, cure or prevent any disease.” FDA Statement page. This supplement-style disclaimer conflicts with the device's own 510(k) clearance for treating post-operative pain and edema and with the site's other claims. source, for the claim These statements have not been evaluated by the Food and...

Has SofPulse been tested?

A gaussmeter or magnetometer reads the field a coil puts out, and an oscilloscope reads its pulse shape and frequency. The field is strongest at the coil surface and falls off fast with distance, so a reading means little without where it was taken. Who paid for the test, and whether the tester earns from links, is printed beside it.

No gaussmeter, magnetometer or oscilloscope reading of a current SofPulse device was found, from the brand or from anyone else.

Certifications claimed, checked

UL, ETL and IEC 60601 are electrical safety marks. ISO 13485 is a quality system. None of them measures the field a device puts out.

SofPulse reviews: what buyers report

Read from public pages on 15 November 2026. A complaint counts as repeated when three or more independent buyers or sites report it.

No Trustpilot, Amazon or BBB rating was found for SofPulse.

Repeated complaints

No complaint was reported by more than one source.

Repeated praise

No praise was repeated across sources.

SofPulse on Reddit

No Reddit thread about SofPulse could be read for this record.

Claims that did not match the record

Each claim is quoted from the seller's own pages and set against the record that contradicts it. FDA, certification, performance and safety claims count against the documentation score. Marketing numbers that do not match are listed but not scored.

Recalls, safety reports and lawsuits

From FDA's recall, adverse event and warning letter records, and court or regulator filings. An adverse event report is an allegation filed with FDA; on its own it does not prove the device caused the event.
TypeDateWhat the record says
FDA adverse event report2007-11-07MW5004370: trial of Ivivi Roma3 (a K070541 model) in a patient with an electronic pacemaker; on the second therapy that day the patient developed shock sensations and multiple electrical pacemaker activity was seen on EKG; unit removed, patient did well, no permanent damage reported. source, for the FDA adverse event report of 2007-11-07
Lawsuit or regulatory action2024-02-02Endonovo Therapeutics filed SEC Form 15 (certification and notice of termination of registration) for its common stock, ending public reporting; 426 holders of record stated. source, for the Lawsuit or regulatory action of 2024-02-02
Lawsuit or regulatory action2023-04-17Endonovo 10-K for 2022 states the company had no employees (about nine contractors) and includes a going concern qualification. source, for the Lawsuit or regulatory action of 2023-04-17

Who sells SofPulse

As found in the seller's own terms and contact pages, BBB and FDA registration, with the source for each.
Company
Endonovo Therapeutics, Inc. (Delaware)
Address on record
6320 Canoga Avenue, 15th Floor, Woodland Hills, CA 91367
Founded
Corporate predecessor incorporated November 2008 in California; SofPulse assets acquired from Rio Grande Neurosciences on December 22, 2017 (Endonovo 10-K filed 2023-04-17)
Owner or parent
Endonovo Therapeutics (OTC ticker ENDV) filed Form 15 on 2024-02-02 to terminate SEC registration. On 2023-12-03 it signed an Asset Purchase Agreement to sell the SofPulse medical-use assets to SofPulse, Inc. (Delaware), closing subject to shareholder approval and other conditions; whether it closed was not found. The 2008 FDA clearance was issued to Ivivi Technologies, Inc.
Customer support
877-796-1456; info@sofpulse.com (site contact page)

Sources: 1, company record2, company record3, company record4, company record5, company record6, company record

SofPulse: common questions

Is SofPulse FDA cleared?

Holds an FDA 510(k) for its device, per openFDA records checked on 15 November 2026. K070541, held by Ivivi Technologies, Inc., Northvale NJ (original applicant), clears IVIVI SOFPULSE, MODELS 912-M10, ROMA3 AND TORINO II for: Adjunctive use in the palliative treatment of post-operative pain and edema in superficial soft tissue. 3 FDA claims the brand makes do not match the record; they are quoted below.

Is SofPulse legit?

SofPulse is sold by Endonovo Therapeutics, Inc. (Delaware). On the archive's documentation rule it scores 2 of 5 (thinly documented), and its customer record is not rated, because there are not enough reviews. 1 of its FDA, certification, performance or safety claims did not match the record. A score measures how well the seller backs up its claims, not whether the therapy works.

What do SofPulse reviews and Reddit threads say?

The research found too few public reviews to summarise. That is stated rather than filled in.

How strong is SofPulse, and what is it FDA cleared for?

SofPulse does not state a field strength with a unit the research could read. FDA cleared IVIVI SOFPULSE, MODELS 912-M10, ROMA3 AND TORINO II for: Adjunctive use in the palliative treatment of post-operative pain and edema in superficial soft tissue. It is sold on a prescription.

Has SofPulse had a recall or safety report?

The records hold 1 FDA adverse event report and 2 lawsuit or regulatory items, each summarised and linked in the safety section. An adverse event report is an allegation filed with FDA; on its own it does not prove the device caused the event.

How these scores work

Documentation score, 5 points

One point each for: a field strength stated with its unit (gauss, tesla, millitesla or microtesla; 1 gauss is 100 microtesla); a pulse frequency stated; an FDA 510(k) clearance or premarket approval that names the seller's own device (registration and listing alone does not earn it); a gaussmeter, magnetometer or oscilloscope reading by someone other than the seller; no FDA, certification, field strength or safety claim contradicted by a record.

5 is fully documented, 4 well documented, 3 partly documented, 0 to 2 thinly documented.

Customer record, 3 points

One point each for: rated 4.0 or better on a review site with 50 or more reviews; a Better Business Bureau grade of A- or better; no single complaint reported by three or more independent buyers or sites. With too few reviews, the record says “not enough reviews to rate” instead of guessing.

Neither score says whether the therapy works. That is what the evidence pages measure.

What could not be verified

  • Whether SofPulse is still sold in the US in 2026 is unclear: the site footer reads 'Copyright 2018-2019', Endonovo stopped SEC reporting in February 2024, and a December 2023 agreement to sell the SofPulse medical assets to SofPulse, Inc. had unknown closing status. FDA listings for Endonovo and ADM Tronics remain active in openFDA.
  • WebSearch budget for the session was exhausted; Reddit, BBB, Amazon, Google reviews, forums and YouTube were not searched; Arctic Shift not queried.
  • Trustpilot: https://www.trustpilot.com/review/sofpulse.com was reported as 404 by the original research; on recheck (2026-11-15) Trustpilot returned 403 (blocked), so profile existence could not be confirmed either way.
  • FDA warning letter and FTC databases not searched; openFDA shows no recalls for Endonovo, Ivivi or SofPulse.
  • No price, warranty or return policy published.
  • No field strength published; no independent lab test found.
  • Transfer of K070541 from Ivivi Technologies to Endonovo not documented in records retrieved.

No seller paid to be listed or rated. For what PEMF has been measured to do in people, read the PEMF evidence.